Herunterladen Inhalt Inhalt Diese Seite drucken

CONDTROL IR-CAM1 Bedienungsanleitung Seite 9

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

EN
THERMAL IMAGER
Automatic image capturing when set temperature limit is exceeded
Enter the menu. Use buttons
,
to select the menu item
and short press the trigger.
By default this function is disabled.
Press and hold the trigger for 1 second to set the temperature
limit and the period of time during which this function will be
active. If temperature limit is exceeded during this period,
the image will be automatically captured and saved in the
gallery.
Use buttons
and
to select adjustment of
temperature limit (Temperature) or period of time, during
which the image will be automatically captured (Duration).
Use buttons
and
adjust each value. It is
possible to set such periods of time – 15 min, 30 min, 45 min,
60 min, 90 min, 120 min.
Short press the trigger to confirm the setting and return to
the menu. Short press button
to exit the menu.
Symbol of this function
If the temperature of the measurement object exceeds the
temperature limit within set period of time, the device will
automatically capture the image and save it in the gallery.
16
IR-CAM1 CONDTROL
User manual
,
,
will appear on the display.
EN
THERMAL IMAGER
To disable this function enter the menu, select menu item
Press and hold the trigger for 1 second. The function will be disabled.
Short press button
to return to menu. Short press
Measurement unit
Enter the menu. Use buttons
,
select the menu item
and press the trigger.
Use buttons
and
to select the measurement
unit:
1) Degrees Celsius – °С
2) Fahrenheit degree – °F
Short press the trigger to confirm the setting and return
to menu. Short press the trigger to confirm the setting and
return to menu. Short press button
Emissivity
All objects emit thermal energy. The volume of energy depends on the surface temperature
and emission degree of the object. The thermal imager measures the intensity of radiation and
uses it to calculate the temperature of the object. Objects with different surfaces at the same
temperature emit different amount of thermal energy. Most of the items and materials, for
example, painted metals, wood, water, leather, fabric have high emissivity (0.9 and more) and
emit more energy than shiny surfaces and unpainted metals with emission degree less than
0.6. Adjustment of emission degree allows the device to take this feature into account and to
minimize the measurement error.