Immunity Test
IEC 60601 Test Level
>95% Dip for 0.5 Cycle
Voltage dips, short interruptions
60% Dip for 5 Cycles
and voltage variations on power
30% Dip for 25 Cycles
supply input lines IEC 61000-4-11
>95% Dip for 5 Seconds
Power Frequency 50/60Hz
30 A/m
Magnetic Field IEC 61000-4-8
Conducted RF
3 Vrms from
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
10 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
For transmitters rated at a maximum output power not listed above, the recommended separation distance D in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note2: T heseguidelinesmaynotapplyinallsituations.Electromagneticpropagationisaffectedbyabsorptionandreflectionfromstructures,objectsandpeople.
Recommended Separation Distances Between Portable and Mobile RF Communications Equipment and this device. This
device and system are NOT Life-Supporting
This device is intended for use in the electromagnetic environment in which radiated disturbances are controlled. The customer or user of this device can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF Communications Equipment and the device as recommended below, according to the maximum
output power of the communications equipment.
Separation distance according to frequency of transmitter M
Rated maximum output power
150 kHz to 80 MHz
of transmitter W
outside ISM bands
D=(0.4)√ P
0.01
0.04
0.1
0.11
1
0.35
10
1.1
100
3.5
For transmitters rated at a maximum output power not listed above, the recommended separation distance D in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note2: T heseguidelinesmaynotapplyinallsituations.Electromagneticpropagationisaffectedbyabsorptionandreflectionfromstructures,objects and people.
A-DV6EB-1
Compliance Level
Electromagnetic Environment - Guidance
>95% Dip for 0.5 Cycle
Mains power quality should be that of a typical commercial or hospital environment. If
60% Dip for 5 Cycles
the user of this device requires continued operation during power mains interruptions, it
is recommended that the device be powered from an uninterruptible power supply or
30% Dip for 25 Cycles
battery.
>95% Dip for 5 Seconds
Powerfrequencymagneticfieldsshouldbethatofatypicallocationinatypical
30 A/m
commercial or hospital environment.
Portable and mobile RF communications equipment should be separated from the
V1 = 3 Vrms
device by no less than the recommended separation distances calculated/listed below:
6 Vrms on ISM & Amateur
D=(0.4)√ P
Bands
D=(0.4)√ P 80 to 800 MHz
D=(0.7)√ P 800 MHz to 2.5 GHz
Where P is the maximum power rating in watts and D is the recommended separation
distance in meters.
Fieldstrengthsfromfixedtransmitters,asdeterminedbyanelectromagnetic
E1 = 10 V/m
site survey, should be less than the compliance levels (V1 and E1).
Interference may occur in the vicinity of equipment marked with the
following symbol:
80 to 800MHz
800 MHz to 2.5 GHz
D=(0.4)√ P
D=(0.7)√ P
0.04
0.07
0.11
0.22
0.35
0.70
1.1
2.2
3.5
7.0
EN - 9