WARNING
NOT allowed for use in hazardous areas!
Equipment used in hazardous areas must be Ex-approved and marked accordingly!
This device is NOT approved for use in hazardous areas!
WARNING
Potential hazard
This device can contain lithium batteries.
Lithium batteries may present a potential hazard if they are abused electrically or mechanically.
Installing/mounting
Wall mounting
Ambient conditions
● Temperature specifications:
Ambient temperature: 0 ... +55 ºC (+32 ... +131 ºF)
Storage: –25 °C ... +70 °C (-13 ... +158 °F)
● Enclosure rating:
IP54 (indoor installation is recommended)
● Environment classes:
EN 1434 class E2 / M2 (electromagnetic / mechanical)
Note
Ensure that the calculator and the sensor cables are installed sufficiently far away from possible sources of electromagnetic
interference (switches, electric motors, fluorescent lamps, etc.).
Ensure that the temperature and flow specifications indicated on the device nameplates / label will not be exceeded.
Note
Temperature sensor pockets
It is recommended to mount the temperature sensors in pockets with welded sleeves.
2
Panel mounting
FUE950 Energy calculator
A5E03652942-002, 04/2014