Herunterladen Inhalt Inhalt Diese Seite drucken
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 19
Note: For instruments with NPT connections, use applicable bonded sealing as shown in
Figure 1
1
Connector

2. Installation

The instrument requires a positive pressure supply, instruments operating in an absolute
range or negative pressure range require a vacuum supply. A vacuum supply should be used
for a fast response for instruments operating near atmospheric pressure. For dual channel
operation two independent pressure and vacuum supplies can be used.
Note: When using two pressure modules (Refer to K0443, Section 4.9, Pressure module
replacement), make sure that:
• The module with the highest pressure rating is fitted to the right hand side Module 1
position as viewed from the rear of the product, refer to (Refer to Figure 1)
• If two modules have the same pressure rating, make sure that the module with the
higher serial number is fitted to the right hand side Module 1 position as viewed from
the rear of the product.
Note: All pneumatic connections must comply with the Pressure Equipment Directive
(PED) or other regional pressure standards.
Note: When connecting the output ports of two pressure modules together make sure
both are either:
• below 70 bar
OR
• between 100 to 210 bar.
© 2008 Baker Hughes, a GE company – All rights reserved.
2
1
2
Figure 1: Pneumatic Connections
Bonded seal
PACE5000/6000 Safety Instructions and User Guide-English | 5

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Pace 5000

Inhaltsverzeichnis