B2.5 Warm-up Cycle

1. The warm up cycle prepares the trophon for operation and will begin automatically when the
machine is powered on or restarted from sleep.
Screen Message
Quick Warm Up
Warming Up
Extended Warm Up
2. Instructions may appear on the screen during this time, which should be followed, such as:
 Close Chamber Door
 Cartridge empty. Replace Cartridge Now?
 Remove Probe from Chamber
3. When warm up is complete, the screen message will read Load Probe and Indicator.
4. The machine is now ready for use.
PART C – ROUTINE USE, MAINTENANCE AND CARE

SECTION C1: LOADING THE DISINFECTANT CARTRIDGE

A trophon disinfectant cartridge needs to be inserted into the trophon EPR before a high level
disinfection cycle can commence. Refer to the trophon disinfectant IFU for detailed instructions on
how to insert and/ or replace trophon disinfectant cartridges.

SECTION C2: ROUTINE HIGH LEVEL DISINFECTION CYCLE

C2.1 Preparing the Probe

Wearing gloves, clean and dry the probe BEFORE commencing the high level disinfection process as
per the probe manufacturer's instructions.

C2.2 Inserting the Chemical Indicator

A trophon Chemical Indicator must be used for each disinfection cycle. No other chemical indicators
are approved for use in the trophon EPR. Each Chemical Indicator may only be used once. Place the
trophon Chemical Indicator into the holder. Refer to the trophon Chemical Indicator IFU. You may
then position the probe in the chamber.

C2.3 Positioning the Probe

1. When the trophon is ready, screen message Load Probe and Indicator will be displayed.
2. Open chamber door.
3. Two clamps hold the probe securely in the chamber. See Figure 10.
The probe has a short sleeve at the back of the handle, covering the electrical cable. This is the
Probe Gland. See Figure 10(a).
Wearing gloves insert the probe correctly in the trophon and ensure the probe gland is held by
the soft clamp. See Figures 10–12.
Ensure the probe is correctly positioned in the chamber. The probe must not contact the
chamber wall and must be positioned above the embossed line at the bottom of the chamber.
L00049 5.0 03/2017
Approximate Warm Up Time
(minutes)
< 2
2–30
> 30
11