Herunterladen Inhalt Inhalt Diese Seite drucken

Getting Started; ConnectingToElectricalPowerSupply; UseWithNebulizerStandardHose; UseWithHeatedHose - DeVilbiss UltraNeb Gebrauchsanweisung

Ultraschallvernebler
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1

2. Getting started

1
2
3
4
5
6
7
8
Aerosol therapy must
only be carried out as
directed by a doctor !
22
Connecting to electrical power supply
Note- M akesurethatthemainsvoltageontheunitsuppliedcorrespondswiththeoperating 
voltageinyourlocation�Ifindoubtpleasecontactanelectricalspecialistoryour
localDeVilbissauthoriseddealer�IfOK,removethesafetylabelcoveringtheEquip-
mentpowersocket�
Connectthemainsplugtotheequipmentpowersocketandtoanearthedmainssocket
outlet�
Use with nebulizer standard hose
1�Inspecthosetocheckfortearsordamagebeforeusing
2�MakesurethatthetubingisproperlyfixedtotheUltraNebconnectors�
3� I ncaseofdamagethetubingmustbereplaced�PleasecontactDeVilbissoryour
authoriseddealer
4� T hetubeshouldonlybedisconnectedatthecuffs�Donotpullhardortwistthetubeas
thetubemaybecomedamaged�
Use with heated hose
1�Useonlythesuppliedhose!
2�Malfunctionsmayariseifothermanufacturersheatedhosesareused!
3� C onnectplugonheatedhosetosocketatbackofnebulizer�Themarkontheplugshould
bepointingupwards�
4� P utheatedsiliconehose(insteadof1200mmhose)totheoutletofthenebulizercham-
berlid,continuousfillingsystemormedicationcup�
Thehoseheatingswitchesonautomaticallywhenultrasonicnebulizerstarts�
Theoperatingtemperatureofmax�37°Cisreachedafterabout2minutes�
Check the operation
Aftercleaning,beforeusesetthemainswitchlocatedontherearpanelto"I"(ON)�
Unitwillstartrunning
Checktheoperationcontrols
- n ebulizer
- a irflow
- t imer
YellowLEDshowswhichleveloffunctionisselected�Push+/-toin/decreasefunctioninsteps
1� I ttakesabout2minutesforheatedhosetowarmup�
2� T otestthealarms,RemovenebulizerchamberfromUltraNeb
3� R edLEDcomesonandalarmsounds
4� T healarmwillsoundfor12secandthenfallsilentfor4sec�Thiswillcontinuefora
periodof2minbeforestopping,theRedLEDwillremainon
5� R eplacethenebulizerchamberbackintoUltraNeb�
6� P ushtheswitchON/OFF(O/I)theRedLEDwithgoout�
7� T heUltraNebisreadyforuseagain�ThegreenLEDislitindicatingstand-bymode�

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis

Fehlerbehebung

loading

Inhaltsverzeichnis