ELECTRONICA-Technologies
The Audiometer 9910 is intended for use in the electromagnetic environment specified below. The customer or the user of
the Audiometer 9910 should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC61000-4-2
Electrical fast
transient/burst
IEC 61000-4-5
Surge
IEC 61000-4-5
Voltage dips,
short
interruptions and
voltage variations
on
power supply
input
lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
Magnetic field
IEC 61000-4-8
is the a.c. mains voltage prior to application of the test level
NOTE : U
T
Audiometer 9910
Rev. January 2017
Guidance and manufacture´s declaration – electromagnetic emission
IEC 60601
Test level
±6 kV contact
±8 kV air
± 2 kV for power supply
lines
± 1 kV for
input/output lines
± 1 kV differential
mode
± 2 kV common
mode
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT) for 5 sec
3A/m
Electromagnetic environment – guidance
Compliance level
±6 kV contact
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
±8 kV air
material, the relative humidity should be at
least 30 %.
± 2 kV for power
supply lines
Mains power quality should be
that of a typical commercial or
± 1 kV for
hospital environment
input/output lines
± 1 kV differential
mode
Mains power quality should be
that of a typical commercial or
± 2 kV common
hospital environment
mode
5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
Mains power quality should be that of a
(60 % dip in UT)
typical commercial or hospital environment.
for 5 cycles
If the user of the Audiometer 9910 requires
continued operation during power mains
70 % UT
interruptions, it is recommended that the
(30 % dip in UT)
for 25 cycles
Audiometer 9910 be powered from an
uninterruptible power supply or a battery.
<5 % UT
(>95 % dip in UT)
for 5 sec
Power frequency magnetic fields
should be at levels
3A/m
characteristics of a typical
location in a typical commercial
or hospital environment
.
Version EN1_H
Series 9000 AUDIOMETERS
Page 30