Herunterladen Inhalt Inhalt Diese Seite drucken
Inhaltsverzeichnis

Werbung

Connection capability: Wireless Controller to the VOS 60 control station:
micro B
Connection capability: VOS 60 control station to the computer:
micro B

Error codes

The fault is shown by an error message in the display (C) as following if the error occurs, e.g. Error 4.
Proceed as follows in such cases:
 Switch the device switch (A) off.
 Remove the stirrer tool and remove the instrument from the assembly.
 Reduce the speed and switch on (instrument switch (A)) the device without the stirrer tool.
Error
Cause
Error 2
Motor current sensor fault
Error 3
Temperature inside instrument is too high motor off - Switch off the instrument and allow it to cool down
Error 4
Motor blockage or overload
Error 8
Speed sensor fault or overload
Error 21 Safety relay fault
If the actions described fails to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the service department
- Send the instrument for repair, including a short description of the fault.
micro A
or
micro B
or
micro B
USB micro B
9-pin RS 232
USB micro B
USB micro B
9-pin RS 232
USB micro B
Effect
Solution
motor off - Switch off the instrument
motor off - Switch off the instrument
- Decrease the load on the motor and restart again
motor off - Switch off the instrument
motor off - Switch off the instrument
24
USB micro A
9-pin RS 232
USB A
USB A
9-pin RS 232
USB A

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis