4.2
Calibrating Sartocheck
switch-on
n
SERVICE ?
forward
>
n
#
n
USER ID
>
0
n
7
n
* SERVICE *
forward
>
L
n
n
L
n
CALIBRATION ?
forward
>
Start
TEMPERATURE
24.3°C
n
L
Appendix
40
®
mini (Continued)
At the end, the calibration certificate is generated.
t
y The customer receives the original calibration certificate along with the unit.
n
PRESSURE
0mbar
n
L
n
TARGET VALUE:
Calibration: *
n
*
TARGET VALUE:>3800mb
Calibration: *
n
3
n
8
n
5
n
0
n
*
n
n
PRESSURE
>3850mbar
n
L
n
CALIBRATION ?
forward
Pressure:
the actual pressure is displayed. The pressure
is measured with the old calibration.
>
0mb
Attention:
Do not connect Sartocheck mini to the
pressure supply
Attention:
Connect Sartocheck mini and reference unit
now to pressure supply (ca. 3800 mbar)
Enter the pressure that is displayed on the
reference unit and confirm with the button *
Calibration process is finished!
Sartocheck mini shows the measured pressure.
Create certificate.
>