Made in Germany
- There is no need to turn the meter into different directions like
for the E-field, instead check all three orientations from time to
time as shown in the following pictures.
- In practice it is usually sufficient to "rotate" the meter out of your
wrist until you have the position / direction of the highest
reading (fourth picture below). In this direction the meter shows
the so-called "resulting" field strength resp. magnetic flux
density.
Please note:
Quick movements induce short peaks of pseudo readings that
have nothing to do with actual fields (due to the earth's static
magnetic field)
Let the display settle for 2 seconds after every change of di-
rection.
Recommended Exposure Limit AC Magnetic Field:
Below 200 nT, preferably below 20 nT
(Magnetic flux density at 50/60 Hz).
(Conversion nT to mG (Milligauss): 200 nT = 2 mG)
8
= Resultant
magnetic flux density!
© Gigahertz Solutions GmbH