Braun Thermoscan Pro 3000 Model - WelchAllyn 9600 Plus Gebrauchsanleitung

Kalibriertester

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Directions for Use

Braun ThermoScan Pro 3000 Model

Caution Store thermometers for testing in the same
room as the 9600 Plus Calibration Tester for
approximately 30 minutes prior to testing to the
thermometer to reach ambient room temperature.
Caution The ambient temperature must be stable and
within the range of 18.3°C (65.0°F) – 26.7°C (80.0°F).
Caution Prolonged body contact with the thermometer
may impact measurements. When instructed to wait
before performing the next step place the thermometer
on the work surface.
Figure 7. 9600 Plus Calibration Tester with Braun ThermoScan Pro 3000
1. Clean the probe tip with a cotton swab slightly moistened
with isopropyl alcohol, remove excess alcohol with a clean
cotton swab, and let air dry for 5 minutes. Do not use any
chemical other than alcohol to clean the probe window.
Chapter 3 Operation
23