9 Repairs

Should the FEC 22 electronic insert need to be repaired by Endress+Hauser, please
send it with a note containing the following information:
• An exact description of the application for which it was used.
• The chemical and physical properties of the product measured.
• A short description of the fault.
Special precautions must be observed when sending the probe and its FEC 22 for
repair:
• Remove all traces of product. This is particularly important if the product can impair
health, i.e. corrosive, poisonous, carcinogenic, radioactive etc..
• If the last traces of dangerous products cannot be removed, e.g. product has
penetrated into fissures or diffused into plastic parts, we kindly ask you not to send
the probe for repair.

10 Waste Disposal

All sales and transport packaging from Endress+Hauser is produced in conformance
to the regulations governing packing for reuse and recycling.
For a small charge, Endress+Hauser will accept and recycle any instruments
manufactured in its own E+H production program. These will then be disposed of
according to the German regulations covering the disposal of electronics. Delivery to
Endress+Hauser, Hauptstr 1,
79689 Maulburg, Germany.

11 Technical Data

General specifications
Operation and system design
Input
Output
Endress+Hauser
Manufacturer
Designation
Measuring principle
Signal processing
Electrical isolation
Measured variable
Range of capacitance
Operating frequency
AC version
DC version
Endress+Hauser GmbH+Co.
FEC 22 electronic insert
Capacitive
AC version: 2 relays connected in parallel (DPDT)
DC version: Load switching via transistor
Between measuring loop and power supply loop
Level (limit)
For empty detection (uncovered probe) 10 pF...350 pF
500 kHz
Two change-over contacts (DPDT), to withstand load:
- AC 253 V, 6 A max., P~max. 1500 VA, cos ϕ = 1
- DC: 30 V, 6 A max.
125 V, 0.2 A max.
Consumption approx. 18 mA at 24 V:
- Load current up to 350 mA continuous, max. 55 V, with overload and
reverse current protection
- Residual current when blocked less than 10 µA
9 Repairs
Electronic insert
Probe
Packaging
Instruments