Do not use on alternate and continuous voltages > 1000V
•
Indoor use in environments with a maximum pollution level of 2 (EN
50419:2006), temperature of -10 °C to +50 °C and relative humidity below 90%.
•
Respect the value and type of the fuses to avoid damaging the instrument and
canceling the warranty.
•
Fuse: 250V/0.5A quick-acting glass tube, 5x20mm.
•
Use accessories corresponding to safety standards (EN 61010-1:2001) with
1000V min. voltage ratings and surge category II.
•
Before any measurement, ensure correct positioning of the leads on the
ammeter. When the value range of a measurement is not known, begin at the
highest range, and then gradually reduce it until the appropriate range is
achieved: the reading should preferably be in the upper 2/3 of the full scale
deflection.
•
The leads must be disconnected to open the lower half of the meter case.
•
Never connect to the circuit to be measured if the casing is not properly closed.
Features
•
Analog mirror scale
•
Robust Plastic housing
•
DC and AC voltage measurements
•
No batteries required
Accessories
•
Operation manual