Do not use on alternate and continuous voltages > 1000V
Indoor use in environments with a maximum pollution level of 2 (EN
50419:2006), temperature of -10 °C to +50 °C and relative humidity below 90%.
Respect the value and type of the fuses to avoid damaging the instrument and
canceling the warranty.
Fuse: 250V/0.5A quick-acting glass tube, 5x20mm.
Use accessories corresponding to safety standards (EN 61010-1:2001) with
1000V min. voltage ratings and surge category II.
Before any measurement, ensure correct positioning of the leads on the
ammeter. When the value range of a measurement is not known, begin at the
highest range, and then gradually reduce it until the appropriate range is
achieved: the reading should preferably be in the upper 2/3 of the full scale
deflection.
The leads must be disconnected to open the lower half of the meter case.
Never connect to the circuit to be measured if the casing is not properly closed.
Features
Analog mirror scale
Robust Plastic housing
DC and AC voltage measurements
No batteries required
Accessories
Operation manual