Herunterladen Inhalt Inhalt Diese Seite drucken

Konica Minolta PULSOX-300 Bedienungsanleitung Seite 21

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
Error Messages
• "H" blinks.
• Measurement value and
"A" flash alternately.
• "-O-" blinks.
• "-U-" blinks.
Cause
• The pulse is too strong for
correct measurement.
• Motion artifact
The measurement value
is displayed; however,
the accuracy of the
measurement values
cannot be guaranteed.
• The pulse rate exceeds the
upper limit (230 bpm) of
the measurement range.
• The pulse rate belows the
lower limit (30 bpm) of the
measurement range.
Solution
• Attach the probe to the body
correctly. If this message is
displayed when the unit is attached
to the patient correctly, the
measurement conditions are
regarded as blood circulation or
other circulatory problems. If this
occurs, search for a location where
more stable pulse signals can be
obtained.
• If this message reappears even
though the above points have been
checked, the probe may be
malfunctioning. Use another probe.
• Attach the probe to the patient
properly. If this message still
reappears even though the probe
is attached properly, keep the
measuring point as stationary as
possible or attach the probe to
another part of the body where
body movement is relatively low.
• Measurement cannot be performed
when the upper limit of the
measurement range is exceeded.
• Measurement cannot be performed
when the lower limit of the
measurement range is not reached.
E 7

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis