Guidance and manufacture´s declaration – electromagnetic immunity
The USB-300 is intended for use in the electromagnetic environment specified below. The
customer or the user of the
USB-300 should assure that it is used in such an environment.
Immunity
IEC
test
60601
Test
level
Conducted
3 Vrms
RF
150 kHz
IEC 61000-
to 80
4-6
MHz
Radiated RF
3 V/m
IEC 61000-
80 MHz
4-3
to 2,5
GHz
AudioConsole 3
®
USB
Electromagnetic environment –guidance
Compliance
level
Portable and mobile RF communications
equipment should be used no closer to any part
of the USB-300, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency
of the transmitter
3 Vrms
Recommended separation distance
=
d
3 V/m
=
d
=
d
Where P is the maximum output power rating
of the transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in meters
(m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
a
should be less than the compliance level in
each frequency range
Interference may occur in the vicinity of
equipment marked with the following symbol:
1,17
P
1,17
P
80
MHz
2,33
P
800
MHz
b
to
800
MHz
to
2,5
MHz
74