Appendixes
IEC 61000-4-8
Electromagnetic immunity
The Steam sterilizer is intended for use in the electromagnetic environment specified below. The
customer or the user of the Steam sterilizer should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2,5 GHz
Compliance level
3 Vrms
150 kHz to 80 MHz
3 Vrms
3 V/m
80 MHz to 2,5 GHz
3 V/m
environmen
t.
Electromagnetic
environment
guidance
Portable
and
mobile
communications equipment should
be used no closer to any part of the
Steam sterilizer including cables,
than the recommended separation
distance
calculated
equation
applicable
frequency of the transmitter.
d = 1,2 √
d = 1,2 √ 80 MHz to 800 MHz
d = 2,3 √ 800 MHz to 2,5 GHz
where P is the maximum output
power rating of the transmitter in
watts
(W)
according
transmitter manufacturer and d is
the
recommended
distance in metres
(m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic
site
Seite 37
-
RF
from
the
to
the
to
the
separation
survey,a