Error symptom
Instrument displays
0.0 mg/l or 0 % Sat O 2
(sensor in air)
Sensor cannot be calibrated
Sensor cannot be calibrated
even after exchange of
electrolyte and membrane
head
Leakage is displayed
Wrong temp. indication
Mechanical damage of the
sensor
Instrument displays OFL
24

Possible errors

Possible cause
No connection
between meter and
sensor
Defective cable
Contaminated
exchange cartridge
Contaminated
electrode resp.
sensor poisoned
Membrane head
hasn't been really
tightned
Hole in the
membrane
Defective
temperature sensor
Display range
exceeded
Short circuit between
gold cathode and
lead anode
Elimination
Control connection
between meter and
sensor
Send sensor to WTW
Refill sensor acc. to
operation manual, wait
60 min. and recalibrate
Cleaning of electrode
Screw tight membrane
head
Exchange membrane
head acc. to operation
manual
Send sensor to WTW
Send sensor to WTW
Regenerate sensor;
if meter still displays
OFL send sensor to
WTW