5

Using this Manual

Conventions in this manual

SOP´s in this manual

This manual refers to the Smartline Column Oven 4050.
Arrows like this:
,
is asked to press the corresponding arrow keys. The operation of arrow
keys is defined as follows:
Cursor up: ▲
Cursor left: ◄
Cursor down: ▼
Important Hints are marked by the marginal hand symbol.
Special warnings are indicated by the marginal warning sign and
printed in bold letters.
The marginal lamp symbol indicates helpful advice.
The Standard Operating Procedures (SOP) provided with this manual
offer a convenient way of structuring complex tasks in the operation of
your Column Oven 4050. They include step-by-step instructions leading
the user through all routine tasks during operation. They can be used for
documentation purposes and be copied, applied signed, and filed in order
to document the performance of the instrument.
Please operate the instrument and all accessories according to
instructions and SOP´s in this manual. This ensures accurate results
and prolongs the longevity of your equipment.
SOP 1
SOP 2
SOP 3
SOP 4
SOP 7
Please Note the Smartline Column Oven 4050 can reach a maximum
temperature of 145°C. Normal silica gel containing HPLC columns are
not allowed to be operated above 90°C. Please set the allowed
maximum temperature for your columns in use. Also observe the
temperature durability of all the other components that are used.