In order to clean the nebuliser, the connection tubing and the
accessories, follow the notes in the respective instructions for
use.

4.2 Filter change

In normal operating conditions, the compressor filter must be
replaced after every 200 operating hours, but at least once a
year.
Also please check the filter at regular intervals (after every 10
inhalations). If it is soiled (grey or brown colour) or blocked, re­
place it. The filter must also be replaced with a new one if it
has become damp.
The filter cannot be cleaned and then reused!
NOTICE
Only use air filters that are intended by PARI for use with your
compressor, otherwise your compressor might be damaged
or it may not be possible to ensure adequate treatment.
To replace the filter, proceed as follows:
• Place the enclosed filter changer
(11) on the filter (7).
Twist and the filter will slide out of
the compressor housing without
difficulty.
• Position the new air filter with the hole in front and press it
as far as possible against the right, cross­shaped adapter.