The device must be cleaned first, as described in "Cleaning the Enclosure", before using a
general disinfecting agent.
Cleaning the Enclosure
WARNING Always power off the system before cleaning. An electrical shock could occur if the
system is powered on and liquid is spilled into it.
To clean the system's enclosure, use a clean, lint-free cloth moistened with either
warm water and mild soap, a diluted non-caustic detergent, or one of the following
approved cleaning agents:
Ammonia: Dilution of Ammonia <3%
Alcohol: Ethanol 70%, Isopropanol 70%.
• Over time, ink overspray might gather at the base of the device. The device uses a vacuum
system to gather most of this ink on a series of saturation pads.
• Eventually, these pads might need to be replaced. Contact Codonics Technical Support to
determine if pad replacement is necessary.
• If ink has gotten onto the system's enclosure, it can be cleaned with an ammonia based
window cleaner and a lint-free cloth.
• If scanning barcodes is inconsistent or the device is having difficulty scanning, clean the
scanner's glass window.
Disinfecting the Enclosure
Recommended disinfecting agents include:
1 part household bleach and 5 parts water solution
A-456-N
Virex II 256
PDI Sani-Cloth
WARNING Codonics makes no claims regarding the efficacy of the listed chemicals or methods as
a means of controlling infection. Consult your hospital's infection control officer or
epidemiologist.
Disinfecting Precautions
To avoid damage to the device, observe the following general precautions for disinfecting the
device:
• Do not use Povodine, Sagrotan, or Mucocit disinfecting agents or strong solvents (for
example, acetone).
®