Major Dental Equipment Utility Specifications
Ritter Leadex 70 DC Intra Oral X Ray
Electrical
1.
P
POWER
2
R
REMOTE
3
S
STRUCTURAL Where required, ensure wall mounting point for x-ray is
REINFORCE
Ritter Dental
Unit 29. 55-59 Norman Street
Peakhurst NSW 2210
PO Box 30 Mortdale NSW 2223
supply GPO minimum 240v/10a power to
marked location area (refer x- ray mounting template),
Leave 500mm tail as requested.
Power Supply must be installed in accordance with
AS/NZS 3003.2011 for Body Protection with R.C.D
(Residual Current Device)
Check x ray unit for Electrical circuit requirements.
Should be wired to same REMOTE isolation on/off switch
as dental chair system.
for remote control connection, supply 3 core
SWITCHING flex cable from x-ray location to selected
area on wall. leave tail of 600mm each end.
reinforced with minimum 700x700x32mm board
(spanning/mounted to internal wall studs) at location for
mounting of x-ray unit.
Bottom of board to be 110cm above floor level
Reinforcement of wall should be made internally to wall
prior to Gyprock sheeting – ensure no gap between
reinforcement board and Gyprock sheeting.
Refer Manufacturer's template for exact location for
reinforcement of wall for x-ray in relation to floor and
dental system.