Herunterladen Inhalt Inhalt Diese Seite drucken

Safety Instructions; Operating Instructions En - IKA IKATRON ETS-D5 Betriebsanleitung

Vorschau ausblenden Andere Handbücher für IKATRON ETS-D5:
Inhaltsverzeichnis

Werbung

Contens
EN

Safety instructions

Correct use
Unpacking
Factory settings
Installation
Operating modes
Setting the HI-TEMP
Adjusting the set-point temperature
Replacing extension cable and sensor
Error messages
Accessories
Technical data
Safety instructions
Caution! The supply voltage (Limited Power Supply) for the ETS-
D5 must be provided using an IEC 61558 compliant transformer
with double or reinforced insulation. Alternatively, laboratory
equipment that meets the requirements of IEC 61010 or NEC
Class 2 may be used.
Do not touch the temperature sensor while measurements are
being taken. This will prevent incorrect results.
Always use the extension cable H 70 when the media being proces-
sed produces vapour. This ensures that the control unit does not
come into contact with the vapour.
Note: This equipment must only be used with media that have a
flash point higher than the temperature safety limit set for the
magnet agitator and/or heating system used.
The equipment must not be operated in areas where there is a risk
of explosion.
Only use glass encapsulated temperature sensors for electrolysis
ETSD5 0306
procedures.
The stainless steel temperature sensor must not be used with
Page
aggressive media such as acids, caustic solutions or distilled
water, due to the risk of corrosion. The H66 glass sensor should
9
be used in such cases.
10
10
Please ensure that the spiral cable does not come into contact
with the heating plate.
10
10
Take care when touching the temperature sensor!
11
Burn hazard! The temperature sensor may heat up to 450°C
11
when inserted in the media!
11
Please observe all accident prevention regulations applicable to
11
the activity and the work station. Personal protective equipment
12
should be worn when operating the equipment.
12
Danger:
12
The user must check that all conditions have been met for the safe
and correct operation of the equipment before operating the IKA-
TRON
ETS-D5 with a heater that does have a DIN 12878 compli-
®
ant connector, and therefore does not provide cut-out proection.
(The cut-out protection is a mechanism that ensures that the elec-
trical heating circuit is switched off should the contact thermome-
ter (ETS-D5) be broken or if the connector is removed.).
H+P magnetic agitator systems (no break fuse protection
included)
When the supply current for the ETS-D5 is interrupted, the elec-
tronics for the heatable magnetic agitator system will take over
the temperature control. This means that the magnetic agitator
system will heat up to the temperature set (e.g. 300 °C). Readjust
the set-point temperature for the magnetic agitator system or
switch the heating off.
Danger when using flammable media with a low boiling
temperature.
Corning Magnetic Agitator PC-400, PC-420, PC-600, PC-620
When using the ETS-D5 in conjunction with a Corning Magnetic
Agitator it is absolutely essential that the rotary "heat" dial on the
magnetic agitator is set to the "O" (off) position. The ETS-D5 will
If th
pos
re s
War
9

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis