Hameg Instruments HM 8133 Bedienungsanleitung Seite 16

Vorschau ausblenden Andere Handbücher für HM 8133:
Inhaltsverzeichnis
Verfügbare Sprachen

Verfügbare Sprachen

11)?
STA
OPI/OPO:
NMO
AMI/AM2:
*
FMI/FM2:
FN{D?
FRQ
DBM
STO/RCL:
Building a string
A string may contain
separators and termination signs. Commands contain 3
symbols and have the ability to question instrument
messages if they are followed by the ?-symbol. Space-
bar, semi-colon,
or commas may be used as separators.
Data for Frequency, Modulation Depth, or Level can be
specified by either fixed point numbers
notation. Units do not need to be specified; they are
supplied automatically in the information transmitted by
the HM8133.
For example:
FRQ;12.345E+3
FRQ:12345
FRQ:1.2345E+4
All of which result in a frequency of 12,345 Hz.
The STA command provides information on the status of
the following
functions:
Frequency
internal/externat
modulation type (AMO/AMI/...).
OPI
RFI
AMO
OPO
RFX
AMI
FMO
FMI
NM()—
Syntax
errors as well
instrument's specifications are signalled by a "beep" as
well as by appearing on the display.
Software
support
Instrument
drivers
for
LabWindows and LabView (National Instruments) are
available on request at Hameg or your dealer.
Alignment
The complete alignment of the HM 8133 requires a
automated test set-up. For this reason it is recommended
to return the instrument to the factory for alignment.
Type of instrument
Status
Output on]0ff
Modulation
off
Amplitude-Modulation
400/1 OOOHz
read ampl. deviation
Frequency modulation
400/1000Hz
Read frequency deviation
AMfFM
external
Reference frequency
internal/external
Frequency
Level
in dB
Level in mVJuV
Set-up storerecall
up to 64 symbols,
or scientific
Output
(OPI/OPO),
(RFI/RFX),
Modulation
mode
Reference frequency
internal/external
Output on/off
as input
values
outside
the
measurement
including
Reference
and
the
of the
software
Inhaltsverzeichnis
loading

Inhaltsverzeichnis