RF emissions
CISPR 11
Harmonic
emissions
IEC61000-3-2
Voltage
fluctuations/flicke
r emissions
IEC61000-3-3
Table 2
Guidance and manufacturer's declaration-electromagnetic
emission
The Pulse Oximeter is intended for use in the electromagnetic
environment specified below. The customer or the user of the
Pulse Oximeter should assure that it is used in such an
environment.
Immunity test
Electrostatic
discharge(ESD)
IEC61000-4-2
Electrical fast
transient/
burst
IEC61000-4-4
Surge
Class B
N/A
N/A
IEC60601
test level
±8 kV
contact
±15kV air
±2kV for
power
Supply lines
±1 kV for
input/output
lines
±1kV line (s) N/A
The Pulse Oximeter suitable
for use in all establishments,
including domestic
establishments and those
directly network that supplies
buildings used for domestic
purposes.
Compliance
level
±8 kV
contact
±15kV air
N/A
18
Electromagnetic
environment -
guidance
Floors should be
wood, concrete
or ceramic tile. if
floors are
covered with
synthetic
material, the
relative humidity
should be at
least 30%
N/A
N/A