1.1.1 Version 1 - Software EAE1163, CPU Board Part Number W3T355622
To update the firmware, unpack the zip file and copy the two files "*.SREC" and "Boot-
load.ini" to the USB memory stick. Do not use subdirectories!
1
De-energize the electronic module.
2
Remove the housing cover of the electronic module.
3
Plug the USB memory stick into the USB socket (A).
4
Reconnect the voltage supply.
5
Using an insulated screwdriver, briefly press the Update button (D) on the mother-
board.
The firmware update takes approximately 1 to 2 minutes. The status LED (C) flashes
green while the firmware update is in progress. The firmware update is complete when
the green LED no longer flashes. The device reboots.
6
The USB memory stick must now be removed.
7
Fit housing cover.
8
All settings must now be entered again (Chapter 5.16).
9
Calibrate the sensors.
1.1.2 Version 2 - Software EAE1192, CPU Board Part Number W3T572231
To update the firmware, unpack the zip file and copy the two files "*.bin" and "update.ini"
to the USB memory stick. Do not use subdirectories!
The firmware update is not carried out if the same version is already installed on the device.
1
De-energize the electronic module.
2
Remove the housing cover of the electronic module.
3
Plug the USB memory stick into the USB socket (A).
4
Reconnect the voltage supply.
5
The firmware update starts automatically. The status LED (C) briefly flashes slowly in
green at first, in red during the deletion process, then faster in green (firmware update
in progress). The Update button (D) is without function.
The firmware update takes approximately 1 to 2 minutes. When the firmware update is
complete, the device reboots.
6
Remove the USB memory stick after the restart.
7
Fit housing cover.
8
All settings must now be entered again (Chapter 5.16).
9
Calibrate the sensors.