occured
and the cutout must reset before continuing.
To reset the cutout press the red
button on the base of the calibrator.
Repeated operation of the cutout indicates a serious fault: you may need to return
the unit to your supplier for repair.
CL-350A Calibrator OPERATOR
'S MANUA
L
27
Before cleaning your unit ALWAYS disconnect from the power supply and allow to
cool below 50 " C.
Your calibrator can be cleaned by wiping with a damp soapy cloth. Care should
be exercised to prevent water from running inside the unit. Do not use abrasive
cleaners.
The controller is factory preset. See the Controller book for further details.
For safety reasons, most of the controller functions are locked. Please contact your
dealer for further information
THE CONTROLLER MUST NOT BE SET TO CONTROL ABOVE 350°C OR DAMAGE TO THE
UNIT MAY RESULT.
INfORMATlON
3.
If the over-temperature cutout light comes on on the front panel, a fault has
ADDlTlONAL