class 2. The instrument may be used in any position, however,
suffi cient ventilation must be assured as convection cooling is
used. For continuous operation prefer a horizontal or slightly
upward position using the feet.
Do not cover either the holes of the case nor the
cooling fi ns.
Specifi cations with tolerances are valid after a 30 minute warm-
up period and at 23 degrees C. Specifi cations without tolerances
STOP
are typical values of an average instrument.

Warranty and Repair

HAMEG instruments are subject to a strict quality control. All
instruments are burned in for 10 hrs prior to shipment. By in-
termittent operation almost all early failures are detected. After
burn-in a thorough test of all functions and of quality is run, all
specifi cations and operating modes are checked.
In case of reclamations during the two years warranty peri-
od please contact the dealer from whom you purchased your
HAMEG instrument. Customers from the Federal Republic of
Germany may directly contact HAMEG for warranty processing
in order to speed up the procedure.
The proceeding of repairs during the warranty period is subject
to our terms of warranty which are available on our web-site
(http://www.hameg.com). Even after expiry of the warranty
period please do not hesitate to contact our HAMEG customer
service for repairs and spare parts.
Return Material Authorization (RMA):
Before sending back your instrument to HAMEG do apply for
a RMA number either by fax or on the Internet: http://www.
hameg.de.
If you do not have suitable packaging for the instrument on
hand please contact the HAMAG sales department (Tel.: +49
(0) 6182/800 300, E-mail: vertrieb@hameg.de) to order an
empty original cardboard box.

Maintenance

The instrument does not require any maintenance. Dirt may
be removed by a soft moist cloth, if necessary adding a mild
detergent. (Water and 1 %.) Grease may be removed with ben-
zine (petrol ether). Displays and windows may only be cleaned
with a moist cloth.
Do not use alcohol, solvents or paste. Under no cir-
cumstances any fl uid should be allowed to get into
the instrument. If other cleaning fl uids are used
damage to the lacquered or plastic surfaces is pos-
STOP
STOP
sible.

Mains voltage

A main voltage of 115V and 230V can be chosen. Please check
whether the mains voltage used corresponds with the volta-
ge indicated by the mains voltage selector on the rear panel.
If not, the voltage has to be changed. In this case the line fuse
has to be changed, too.
Please note:
After changing the main voltage, the line fuse has
to be changed. Otherwise the instrument may be
destroChanging the line fuse

Changing the line fuse

The fuses are accessible from the outside and contained in
the line voltage connector housing. Before changing a fuse
disconnect the instrument from the line, the line cord must
be removed. Check fuse holder and line cord for any damages.
Use a screw driver to loosen the fuse holder screw counter-
clockwise while pressing the top of the fuse holder down. The
top holding the fuse will then come off. Exchange the defective
fuse against a correct new one.
It is forbidden to repair defective fuses or to bridge them by any
means. Any damage caused this way will void the warranty.
MODULATION
TRIGGER
R
L
IN / OUT
INTERFACE
Types of fuses:
Size 5 x 20 mm; 250V~,
IEC 60127-2/5
EN 60127-2/5
Line voltage
Correct fuse type
230 V
2 x 1.6 A slow blow
115 V
2 x 3.2 A slow blow
I m p o r t a n t h i n t s
VOLTAGE
SELECTOR
230 V / 115 V
F1
F2
Instruments
D-63533 Mainhausen
115/230 V ±10% / 50-60 Hz
Watts (max.): 160
HM8142-2
Fuse: IEC127 - III
PROGRAMMABLE
115 V: 2 x 3,2 A / 230 V: 2 x 1,6 A
POWER SUPPLY
5 x 20 mm slow blow,
!
Made in Germany
träge, temporisé, lento
Subject to change without notice