Herunterladen Inhalt Inhalt Diese Seite drucken

PeakTech 2860 Bedienungsanleitung Seite 21

2,7 ghz frequenzzähler
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Considering:
* When push the HOLD button at once, then the display will show "- - - HOLD
- - -" & holding values alternately.
* It will release the hold function if push the HOLD button at once again.
5.3. Relative Measurement
1. During the measurement, the circuit will memorize the last measured values
if push the "REL. Button" (3-9, Fig. 1) at once, then LCD will show "0" & a
"REL" marker appears on the right down corner.
2. The new measured frequency values will deduct above memorized "last
measured values" automatically.
3. It will release the Relative Measurement function if push the REL. button at
once again, at same time the "REL" marker will disappear.
Considering:
When marking the "Data Hold" & "Data Record" measurement, the Relative
function is prohibited.
5.4. Data Record (Max., Min., Average reading)
1. The DATA-RECORD function displays the maximum, minimum and average
readings. To start the DATA RECORD function, press the RECORD button
once. An "R.C." marker should appear on the top right corner of the display.
2. Push the RECALL button once and "- - -HI- - -" should appear on the display
followed in about a second by the maximum reading. The "R.C." marker will
be flashing.
3. Push the RECALL button again and "- - -Lo- - -" should appear on the
display followed by the minimum reading.
4. Push the RECALL button again and "- - -A- - -" should appear on the display
followed by the average reading.
The average reading will be continually updated every ten samples.
5. Push the RECALL button again will stop the "R.C.", marker from flashing
and normal reading will be.
-20-

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis