INSTRUMENT OPERATION - The instrument is switched on by pressing the ON/OFF
button located on the front of the thermometer. Insert the probe into the substance
being measured. The instrument reading may take a few seconds to stabilise,
depending on the nature of the measurement. The instrument will automatically
turn off after 10 minutes.
HOLD - Press the HOLD button to retain the display reading, press the HOLD button
again to resume taking temperature. Please note: once HOLD is selected the unit will
not automatically turn off.
°C/°F - While powering the unit on, hold the ON/OFF button for five seconds to
change between °C and °F. Once changed, the unit will remain in the selected value.
CALCHECK - To check the calibration of the thermometer press and hold the HOLD
button, turn the instrument ON and the test cap within the unit will display 0.0 °C ±0.1 °C.
BATTERY REPLACEMENT - The battery will need to be replaced when the
instrument displays
'FLAt bAtt' or if the battery empty icon is displayed. Replace the battery with a 3 volt
lithium coin cell (CR2032). The battery is located at the top of the instrument, twist
the cap anti-clockwise and pull to remove.
ERROR MESSAGE - If the thermometer or probe has developed a fault the
instrument will show 'Err'.
EMC/RFI - Instrument performance may be affected temporarily if operated within
a high frequency radio field, such as near a mobile phone, or if subjected to an
electrostatic shock.
GUARANTEE - This instrument carries a two-year guarantee against defects in
either components or workmanship. During this period, products that prove to be
defective will, at the discretion of ETI, be either repaired or replaced without charge.
This guarantee does not apply to probes, where a six-month period is offered. The
product guarantee does not cover damage caused by fair wear and tear, abnormal
storage conditions, incorrect use, accidental misuse, abuse, neglect, misapplication
or modification. Full details of liability are available within ETI's Terms & Conditions
of Sale at etiltd.com/terms. In line with our policy of continuous development, we
reserve the right to amend our product specification without prior notice.
FR - FONCTIONNEMENT DU THERMOMÈTRE
Mettez l'instrument en marche en appuyant sur le bouton ON/OFF situé sur la face
d'avant. Insérer la sonde dans la substance à mesurer. La stabilisation de l'indication
de l'instrument peut prendre quelques secondes, selon la nature de la mesure.
L'instrument s'éteindra automatiquement après 10 minutes.
HOLD - Appuyez sur le bouton HOLD pour conserver la valeur affichée, appuyez
une seconde fois sur le bouton HOLD pour recommencer la prise de la température.
Remarque: Une fois HOLD est sélectionné, le thermomètre ne s'éteint pas
automatiquement.
°C/°F - Lors de l'allumage de l'appareil, maintenez enfoncé le bouton ON-OFF
pendant cinq secondes pour basculer entre °C et °F. Après la modification, l'appareil
reste dans l'unité sélectionnée.
CONTRÔLER LE L'ÉTALONNAGE - Pour vérifier l'étalonnage de l'instrument, presser le
bouton HOLD pendant l'allumage de l'instrument. L'unité procédera alors à un auto-
étalonnage et affichera 0.0 °C. L'exactitude de cette valeur affichée est de ±0.1 °C.
REMPLACEMENT DE LA PILE - La batterie doit être remplacée lorsque 'FLAt bAtt' ou
l'icône batterie vide s'affiche. Remplacez la pile par une pile bouton au lithium de 3
volts (CR2032). La pile est située dans la partie supérieure de l'instrument, tournez le
capuchon dans le sens anti-horaire et tirez dessus pour le retirer.