CALIBRATION PROCEDURES
1. Using a acoustic calibrator
Make the following switch settings.
RANGE
RESPONSE
FUNCT
Insert the microphone carefully into the insertion hole
of the calibrator.
Tum on the switch of calibrator and adjust the
CAL screw of the instrument, until the level dis-
play indicates the desired level. Note: Our pro-
ducts
are
well
Recommended
calibrator
2. Calibration using the intemal oscillator
Make the following switch settings.
RANGE
RESPONSE
FUNCT
Display will show 94.0 ± 1.5dB
SOUND LEVEL METER
Hi
F
A
calibrated
before
shipment.
cycle
is
one
Hi
F
CAL 94dB
ACOUSTIC
CALIBRATOR
TE S
1351
SOUND LEVEL METER
dB
75. 0
RANGE
RESPONSE
FUNCT
Lo
S
A
Hi
C
F
POWER
MAX
CAL 94 dB
OFF
HOLD
Reset
Lo = 35
90dB
Hi = 75
130dB
IEC 651, ANSI S1.4 TYPE 2
ACCURACY: ± 1.5dB
VORGEHENSWEISE DER KALIBRIERUNG
1. Verwendung eines externen Kalibrierungsgerätes
Stellen Sie das Gerät wie folgt ein:
RANGE
RESPONSE
FUNCT
Fügen Sie das Mikrofon vorsichtig in die vorgesehene
Öffnung des Kalibriergerätes.
Schalten Sie den Kalibrator ein und drehen Sie die Ka-
librierungsschraube des Messgerätes bis im Display
der gewünschte Level eingestellt ist.
Hinweis: Unsere Produkte werden sorgfältig kalibriert,
year.
bevor sie unser Haus verlassen. Der empfohlene Kali-
brierungszyklus beträgt 1 Jahr.
2. Verwendung des internen Oszillators
Stellen Sie das Gerät wie folgt ein:
RANGE
RESPONSE
FUNCT
Das Display zeigt 94.0 ± 1.5dB