Verfügbare Sprachen

The repair service may only be performed by staff authorized
and trained by the manufacturer. A modification of the system is
not permitted.
Caution when handling aggressive chemicals
Do not use aggressive chemicals like e.g. strong and weak bases,
strong acids, formaldehyde, acetone, halogenated hydrocarbons
or phenol for cleaning the system and its accessories.
Cleaning
1. Disconnect the system from the power supply before you start
cleaning.
2. Clean all outer parts of the system with a mild detergent and a lint-
free cloth.
3. Wipe off the detergent with Aqua dest.
4. Dry all cleaned parts.
Desinfection
1. Disconnect the system from the power supply before you start
disinfecting.
2. Let the system cool down.
3. Clean the system as described above.
4. Select a disinfection method compliant to the applicable local legal
regulations and directives.
5. Wipe off all outer parts of the system with the disinfectant and a lint-
free cloth.
slit).
Do not perform spray disinfection.
Do only connect the system with the power supply if it is
completely dry.
In case of contamination with aggressive chemicals,
clean the system with a neutral detergent immediately.
Use neither corrosive detergents nor aggressive solvents
or abrasive polishing agents.