After the completion of the aspiration process, the handle should be rinsed to
avoid a blocking of the valve. During rinsing, slowly close and open the valve.
If the valve is blocked, please contact our service department or remove the
little lower knurled screw and try to loosen the valve by means of dissolving
liquid.
Technical Data
Vacuum pump
Aspiration capacity
Final vacuum
Material of pump head
Housing
Dimensions (W x L x H)
Height with bottle 02 l
Height with bottle 04 l
Weight without accessories
Material housing
Power input
Electr. supply
Fuse
Environmental conditions
ambient air temperature
in service (non-condensing)
relative humidity

Connection of additional units

This description refers to the connection of additional units to the systems
AA 02, AA 04 and AA 24
Up to 4 additional units can be connected to the aspiration systems. This can
be done with 4 bottles with one handle each.
AA 02 / 04, AA 24
17 l / min
100 mbar abs.
aluminium
170 x 220 x 155 mm
440 mm
510 mm
5.5 kg
aluminium
100 W
230 V / 50 Hz
2 AT
4 – 40°C
max. 80%
11