EN
the device down during use, please press the "ON/OFF" button for more
than 1 second.
Attention: At the end of treatment, it is normal that a little medication
left in medication cup after automatically shutdown.
Chapter 5 Maintenance, Transport and Storage
5.1 Clean and disinfection
Clean and disinfect the device after each use. If the device doesn't
clean up, the nebulizing will be affected because of drying and
coagulation of medication.
1. Remove medication cup, accessory and batteries from the main unit.
2. Open the cup lid and discard residual medication.
3. Add 75% ethanol solution in medication cup, cover the cup lid, then
leave for at least 10mins; it is available to gently shake it for better
disinfection.
4. Immerse the accessories to be disinfected into a container with
ethanol solution, and lid the container. Use 75% ethanol solution soaking
for 10mins or longer.
5. Discard the disinfectant in medication cup, take accessories out from
the disinfectant; clean the medication cup and accessories with clear-
water repeatedly.
6. Fill medication cup with clear-water, assemble it to the main unit, let
the device work 10mins in order to clean the nebulizing sheet.