Rated conditions
● Electromagnetic
compatibility
Interference emission and
interference immunity
Process medium conditions
● Process medium
temperature
Pressure transmitter construction
Weight
● Pressure transmitter
Material
● Wetted parts materials
Housing
Sensor
Cable
Protective cover
Sealing material
Electric connection
Torque for cable gland nut
made of
Cable box 7MF1572-8AA construction (accessory)
Field of application
Weight
Electric connection
Cable entry
Enclosure material
Vent pipe for atmospheric
pressure
Screw for bearer wire
Rated conditions
● Degree of protection
according to EN 60 529
Anchor clamp 7MF1572-8AB construction (accessory)
Field of application
Weight
Electric connection
36
To EN 61326-1 and EN 61326-2-3
-10 ... +80 °C (-4 ... +176 °F)
Approx. 0.2 kg (0.44 lb)
Stainless steel, mat. no. 1.4404 or AISI 316L
Ceramic AI
O
(96%)
2
3
PE-HD
PPE
FPM, EPDM (for drinking water)
Cable PE-HD: Lengths 2, 5, 10, 15, 20, 30 m
Plastic
2.5 Nm (1.8 ft lb)
For connecting the transmitter cable
0.2 kg (0.44 lb)
2 x 3-way (28 to 18 AWG)
2 x Pg 9
Polycarbonate
IP66
For fastening the transmitter
0.16 kg (0.35 lb)
Galvanized steel, polyamide