Technical Data (5.101.000, 5.102.000, 5.103.000)
Mecanical
Quartz glass heating element:
Ceramic annealing tube:
Ceramic annealing tube (length):
Angular adjustments:
Weight:
Measurements (w x h x d):
Casing:
Heating element:
Annealing tube:
Heating element Insulation:
Electical
Sterilization temperature:
Power supply:
Fuses:
Protection class:
Heating up time:
Sterilization time:
Heat transfer:
Enviromental conditions
Safety features
Residual heat display:
Licences
CE:
EU guidelines:
*tested with inoculation loops, WLD-TEC (Art.-No.: 6.000.373)
Ø 26,6 mm
Ø 20 mm
161 mm
7
1060 g
175 x 129 x 169 mm
stainless steel
quartz glass
ceramic, 60% Al
O
2
3
MgO 19-26%, SiO
72-77%
2
900°C - 950°C (1650°F - 1742°F)
Art. No.: 5.101.000:
220 - 240 V / 50/60 Hz / 225 W
Art. No.: 5.102.000:
112 - 128 V / 50/60 Hz / 225 W
Art. No.: 5.103.000:
95 - 105 V / 50/60 Hz / 225 W
Art. No.: 5.101.000:
1,25 A (F)
Art. No.: 5.102.000:
2,0 A (F)
Art. No.: 5.103.000:
2,5 A (F)
I
700°C / 15 minutes; 800°C / 22 minutes
5 - 7 seconds*
a stainless steel surface, located above, warms up at
motionless air and at an ambient temperature of 20°C;
Distance: 20 cm: 65°C, 30 cm: 40°C, 40 cm: 32°C
Minutes
operating temperature 5°C bis 40°C. Maximum
relative humidity 80 % for temperatures up to 31°C,
decreasing linearly to 50 % relative humidity at 40°C
indicates a hot annealing tube
EN 61326-1, EN 61010-1, EN 61010-2-010
2014/30/EU, 2014/35/EU, 2011/65/EU
EN
27

Werbung

loading

Diese Anleitung auch für:

5.001.0005.002.0005.003.000