Section 3 Installation
CAUTION
Remote Temperature
Sensor (6200 Only)
WARNING
3-4
Ensure all communication and electrical connections are made
prior to starting the unit and that the cords do not come in contact
with any of the plumbing connections or tubing.
For refrigerated units, ensure the line cord from the bath is securely
connected to the rear of the controller.
For all units, ensure the communications cable from the bath is
connected to the RJ45 connectors (similar to Ethernet) on the rear of
the controller.
For all units, connect the supplied line cord to the controller and to a
properly grounded outlet.
Communications cable
from bath
Cable connection required
for proper communication
between the bath and the
circulator.
Line cord from bath
(refrigerated baths only)
Return Flow
6200 USB
Connection
Outlet Flow
Never apply line voltage to the connection.
The optional remote temperature sensor is enabled using the controller,
see Section 4.
Pin
Pin
1
White
7
Red
2, 3
NA
8
NA
4
White
9
Red (4th wire not connected to the control board)
5, 6
NA
4100/5150 Immersion Circulator
6200 Immersion Circulator
I/O Circuit Protector
Line cord connection
Outlet Flow
6200 Remote
Temperature Sensor
1 2 3 4 5
6 7 8 9
Return Flow
Fisher Scientific