Trouble-shooting
92
9.7
Software-/firmware history
Release software (SW)-/firmware (FW) version
The SW-/FW version indicates the device release history: XX.YY.ZZ (example 01.02.01).
XX
Change in the main version.
No longer compatible. Changes to device and Operating Instructions.
YY
Change in the functionality and operation.
Compatible. Changes to Operating Instructions.
ZZ
Debugging and internal modifications.
No changes to Operating Instructions.
SW-/FW-version, date
01.01.00, 09/2002
01.02.00, 12/2002
01.03.00, 09/2004
01.03.01, 04/2005
01.03.03, 12/2006
Operation, documentation
Compatible with:
• HART Communicator DXR275 (as
of OS4.6) with DevRev1, DDRev 1
®
• Readwin
2000 version 1.9.1.1
• Commuwin II (as of version
2.07.01-4)
• AMS (as of version 5.0)
• PDM (as of version 5.1)
Compatible with:
®
• Readwin
2000 version 1.10.1.1
Compatible with:
• HART Communicator DXR275
(from OS4.6) with DevRev 2,
DDRev 1
• HART Communicator DXR375
(from OS1.6) with DevRev 2,
DDRev 1
®
• Readwin
2000 (as of version
1.16.2.0)
• AMS (as of version 5.0)
• PDM (as of version 5.1)
• FieldCare version as of 2.01.00
TMT162
Modifications
Original firmware
Parameter for trimming 4 to 20 mA
loop
• Customer-specific linearization,
sensor matching f. RTD sensors
• Callendar Van-Dusen coefficients
for Pt100
• New sensors:
Pt100 SAMA ( = 0.003923)
Cu10 ( = 0.00427)
Pt200 IEC 751 ( = 0.00385)
Ni120 ( = 0.00672)
Pt50/100 GOST ( = 0.003911)
Cu50/100 GOST ( = 0.004278)
• Adjustable error current value
(between 21.6 and 23 mA)
• Measured value shown on display
with unit %
• Adjustable number of figures
after decimal point
®
New HART
command 231 and
minor bugfixes
Internal SW modifications
Endress+Hauser