Herunterladen Inhalt Inhalt Diese Seite drucken

Important Safety Instructions - Spectro-UV Spectroline MAXIMA ML-3500-Serie Bedienungsanleitung

Uv-a-lampen mdl-mikroentladungslampe
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

LAMP CONTAINS MERCURY
Manage in Accordance with
Disposal Laws
See: www.lamprecycle.org
English
MAXIMA
Micro Discharge Light (MDL) UV-A Lamps
U.S. Patent No. 6,573,665
INTRODUCTION
The MAXIMA
ML-3500 series ultra-high intensity UV lamps feature micro discharge light (MDL)
technology to make all inspections easier, safer and more reliable. The MAXIMA series lamps are
available in three line-voltage models: ML-3500S, with a spot bulb;
ML-3500D, with a spot bulb and a diffusing filter; and ML-3500FL, with a flood bulb for a wider
range. All "M" model MAXIMA lamps operate in both alternating (AC) or direct current (DC). Each
model, ML-3500MS (spot), ML-3500MD (diffusing filter) and ML-3500MFL (flood) includes a 12-
volt, 7 amp/hr rechargeable battery that will operate the lamp for two hours and a battery charger
and carrying case. Each comes equipped with a detachable AC cord for the country of shipment
(120V, 230V, 240V, or 100V).
WARNING:
UV light emitted from this product.
Avoid eye and skin exposure to unshielded products.
FOR PROFESSIONAL EXAMINATION USE ONLY.

IMPORTANT SAFETY INSTRUCTIONS

1. Read all instructions.
2. Care must be taken, as burns can occur from touching hot parts.
3. Do not operate the lamp with a damaged cord or if the lamp has been damaged.
4. Wear UV-absorbing protective eyewear (such as the UVS-30 glasses provided) when operating
this or any UV lamp.
5. The filter lens will become hot during use. Never operate the lamp with the UV filter lens
removed.
6. Never use this equipment in any manner not specified in these instructions because the
protection provided by the equipment may be impaired.
NOTE: Due to the ultra-high UV-A intensity produced by this lamp, subject (penetrants,
metals, plastics) fluorescence may fade after a period of exposure. Therefore, this lamp
should not be left shining on any one area for an extended period.
ML-3500S AC OPERATION
1. Connect the plug to a power source that complies with the electrical requirements specified
on the lamp's label. For best performance and durability, the power source should be
regulated, surge free and noise free. In those regions where mains may experience
fluctuations greater than ± 10%, an overvoltage protection device is available.
2. Press ON the BLACK ballast rocker switch.
ELECTRICAL SPECIFICATIONS—ML-3500S MODELS
MAXIMA
ML-3500S
ML-3500S/F
Model
120V/60Hz/
230V/50Hz/
Power Source
1.00A
0.75A
ML-3500 Series
ML-3500S/FB
ML-3500S/FA
ML-3500/SJ
230V/50Hz/
240V/50Hz/
100V/50-60Hz/
0.75A
0.75A
1.00A
3

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Spectroline maxima ml-3500s

Inhaltsverzeichnis