Battery charging
The battery charges automatically when installed in an approved
Tektronix instrument that is connected to an external power supply. You
can also charge the battery with the Tektronix external battery charger
(TEKCHG-XX). Remove the battery from the charger when charging is
complete.
A lithium-ion battery will self-discharge during non-use. To achieve the
longest operating time, charge your battery before use. If you plan to
store batteries, read the instructions in the Battery storage section of this
document.
Configuration
Typical charging time
One fully discharged battery
Time varies per instrument type
charging in the instrument.
and the instrument's power usage
while charging. See instrument
user manual for charge time.
One or two fully discharged
5 hours
batteries charging in the Tektronix
external battery charger
CAUTION: To avoid damage to the battery, use only the
Tektronix external battery charger or an approved instrument
to charge the battery. Do not connect any other voltage source
to the battery.
For optimum performance, charge the battery completely before using it
for the first time or after prolonged storage.
To discharge the battery completely, continue to run the instrument
on the battery until automatic shutdown occurs. After the instrument
shuts down, either recharge the battery promptly or remove it from the
instrument until you are ready to recharge the battery.
Note: To prolong the life of the battery and to prevent
shutdown, do not operate or charge the battery at high
temperatures. For best results, allow the battery to cool to room
temperature before using or charging the battery.
Do not leave a battery discharged for extended periods. See
the Storing batteries section for information on how to properly
store a battery.
The battery stops charging under the following conditions:
The battery is fully charged.
The internal temperature of the battery exceeds the safe charging
threshold.
A fault condition occurs.
The battery shuts down and disconnects itself from the instrument under
the following conditions:
The battery is fully discharged.
The internal temperature of the battery exceeds the safe discharging
threshold.
A fault condition occurs.
Battery handling
Do not disassemble, crush, drop, or puncture the battery.
Do not short the external contacts on the battery.
Do not expose the battery to fire or water.
Do not expose the battery to temperatures above +60°C (+140°F).
Seek medical advice if a battery or part of it has been swallowed.
Do not put batteries near heat or fire. Do not put in sunlight.
Keep the battery clean and dry. Clean dirty connectors with a dry,
clean cloth.
Batteries contain hazardous chemicals that can cause burns or
explode. If exposure to chemicals occurs, clean with soap and water
and get medical aid. Repair the product before use if the battery
leaks.
Do not attempt to open, modify, reform or repair a battery
that appears to be malfunctioning, or which has been physically
damaged.
Keep the battery away from children and animals.
Avoid exposing the battery to excessive shock or vibration.
Do not use a damaged battery or charger.
If a battery has leaking fluids, do not touch any fluids. Dispose of a
leaking battery. See the Battery recycling section for information on
disposal and recycling.
In case of eye contact with fluid, do not rub eyes. Immediately flush
eyes thoroughly with water for at least 15 minutes, lifting upper
and lower lids, until no evidence of the fluid remains. Seek medical
attention.
Battery storage
Store batteries in a low-humidity environment free of corrosive
gases. Storing batteries in high-humidity environments, or outside
the temperature range, can cause oxidation on the metallic parts,
increased leakage, significant aging and premature failure.
Do not store batteries near heat or fire. Do not store in the sunlight.
Do not keep batteries in a place where the terminals can be shorted
by metal objects (for example, coins, paper clips, or pens).
Do not remove a battery from its original packaging until required for
use.
When possible remove the battery from the equipment when not in
use.
Charge the battery to approximately 50% of capacity at least once
every six months.
Remove the battery and store it separately from the instrument.
Refer to the documentation for your instrument for instructions on
removing and installing the battery.
For optimal battery life, store the battery at temperatures between
+5°C and +25°C (+41°F and +77°F). Higher temperatures reduce
the battery storage life.
Battery replacement
Replace the lithium-ion battery only with a TEKBAT-XX battery.
See the Battery recycling section for information on disposal and
recycling.
Specifications
Characteristic
Description
Capacity, typical
6.7 Ah (97.28 Wh)
Output, nominal
14.52 V
Charge
0°C to +45°C (+32°F to +113°F); varies per
temperature
charging current and battery heat dissipation
characteristics. The actual limit may be lower.
Tektronix recommends using the TEKCHG-XX
external battery charger to charge the TEKBAT-XX
batteries in ambient environments greater than
+30°C.
Discharge
-20°C to +60°C (-4°F to +140°F); varies per
temperature
the discharge current and battery heat dissipation
characteristics. The actual limit may be lower.
Storage
-20°C to +35C (-4°F to +95°F)
temperature
Weight
Battery: approximately 447.4 g (15.78 oz)
Battery (including packaging): approximately 564 g
(19.89 oz)
Copyright © Tektronix. All rights reserved. Licensed software products
are owned by Tektronix or its subsidiaries or suppliers, and are protected
by national copyright laws and international treaty provisions. Tektronix
products are covered by U.S. and foreign patents, issued and pending.
Information in this publication supersedes that in all previously published
material. Specifications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc.
14150 SW Karl Braun Drive
P.O. Box 500
Beaverton, OR 97077
USA
For product information, sales, service, and technical support:
In North America, call 1-800-833-9200.
Worldwide, visit to
www.tek.com
find contacts in your area.
Warranty
For warranty information, go to
www.tek.com/warranty-status-search

Werbung

loading

Diese Anleitung auch für:

Tekbat-01