System Messages; Maintenance And Support - Novasina CaliBox 200 Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Brief instruction to software installation:
1. Put the CD in the PC disk drive and select the
drive directory with the Windows Explorer.
2. Proceed with the installation selecting the No-
vasina_PascalTool WIN.exe data fi le -> fi nish
the SETUP.
3. The PascalTool WIN software is now installed.
Connection of PC/Laptop to the CaliBox 200:
The supplied data cable (DB-9 to Stereo-plug) can
be connected to the stereo socket (Interface), which
can be found above the display.
Operating the PascalTool WIN software :
Start the PascalTool WIN software as soon as the
CaliBox 200 is connected to the PC/laptop and all
devices are switched on.
Brief instruction:
1. Start the software -> main window appears
2. Activate the COM-Ports (is USB interface is
used). The comport can be selected in the menu
Options -> Preferences -> COM Port -> OK.
3. Read out the data -> Read Data, afterwards the
following device information appears (data of
integrated CaliBox 200 sensor)
- Instrument no.:
Pascal-STV Z
- Serial No.:
- dP range:
- Pressure:
(currently measured value)
Remark: In case of error messages please select
another COM-Port.
4. Selection of one of the following registers for the
activation of the required function.
- Measurement:
Sensor calibration by gain
adjustment
pass
fi lter for smoothing of the
actual value
- Analog Output: Scaling of the analogue
output
Selection of the voltage
output range 0/2....10V
Stereo socket
for PascalTool WIN
data cable
XXXXXXX
0...200Pa
XXX.XXPa
Adjustment of low-
Analogue output
calibration by gain and
offset adjustment
- LC Display:
Selection of measurement
value unit and display
contrast
5. Press Write Data in order to save all device
settings.
Remark: The PascalTool WIN software offers also
the possibility to protect the settings by a personal
password. This can be activated by selecting the
function Options -> Change protection code.
The settings can also be printed out. This is done
by the function File -> Print.
PascalTool WIN offers also a help program. This
function can be found in the main menu bar and con-
tains all relevant information and software functions.

9. System messages

AZ OFF
auto-zeroing off
AZ ON
auto-zeroing on
ZERO CAL
Manual zeroing in progress
dP UNDERFLOW Pressure below min. level of
measurement range (<= 5 Pa)
dP OVERFLOW
Pressure above max. level of
measurement range (>200 Pa)
NOVAS SELFCAL Performance of self
calibration (self testing when
device is switched on)

10. Maintenance and support

Maintenance/cleaning:
The device can be wiped off with a damp cloth. Don't
use any aggressive cleaning agent!
Accreditation & recalibration:
In case of a later certifi cation of the device by an
accredited calibration laboratory please contact
you local Novasina distributor. He will support you
accordingly.