temperature shown on the contact-thermometer not be reached then this
is normally due to the fact that the hotplate temperature has been set too
low. Please ensure that the temperature-limiter is also accordingly set at a
higher temperature.
a. Operating with an electronic contact thermometer
An electronic C-Thermometer can be directly connected to the 5-pole
socket at the back of the appliance. The necessary power for the contact
thermometer will be supplied over this socket.
The required temperature should now be adjusted on the contact
thermometer (see the contact thermometer operating instructions).
b. Operating with a mercury contact thermometer
When operating with a mercury contact thermometer, our cable (Order
No.. SK 85) must be connected to the thermometer.
Please make sure that the heating output approximately matches the
heat-requirement, this can be adjusted with the power regulator (turning
knob 10-100%), High surplus heating output always results in post-
heating. The more precise the heating output adjustment is, the more
exact is the temperature tolerance.
c Operating with a temperature cut-out, Timeswitch, etc..
Instead of the bridge between 3 and 5, other contact-giving devices may
be connected that will switch off the heating when the contact has been
opened.
When using a temperature cut-out please make sure that the cut-out has
a thermal contact to the medium being heated
Please note that no external voltage should be used on the contact
socket.
Information: the contact-thermometer socket is fitted with 12 V
(galvanically separated).