5
Installation
Before starting do the following:
Read and understand the "Safety" section.
Do not use damaged equipment.
Note: Use only original parts supplied by the manufacturer.
5.1 External pressure connections
Ref: figure A3 (front cover). Use an applicable method to seal the external pressure connections,
and then tighten to the applicable torque. Maximum torque:
1/8 NPT: 35 Nm (26 lbf.ft)
G1/8: 25 Nm (18.4 lbf.ft)
5.2 Calibrator assembly
Optional item. (DPI 620G/PM 620). See figure B2 (front cover).
Note: To use the DPI 620G calibrator, read the user manuals: K0541 (Druck DPI 620G Advanced
modular calibrator) and K0457 (Druck PV 62XG pneumatic/hydraulic pressure stations).
.
B2
Step
1.
2.
3.
4.
6
Maintenance
Clean the case with a moist, lint-free cloth and a weak detergent. Do not use solvents or abrasive
materials.
Return the instrument to the manufacturer or an approved service agent for all repairs.
Do not dispose of this product as household waste. Use an approved organisation that collects and/or
recycles waste electrical and electronic equipment. For more information, contact one of these:
Customer service department: Druck.com
Local government office.
8
Quick Reference
Procedure
Lower the calibrator into the moulded compartment.
Press on the bottom end of the calibrator until it latches in position.
Attach a pressure module with the correct range and type.
Tighten it until it is hand tight only.