Herunterladen Inhalt Inhalt Diese Seite drucken

Waeco AirCon Service ST-8810 Bedienungsanleitung Seite 23

Infrarot-thermometer
Vorschau ausblenden Andere Handbücher für AirCon Service ST-8810:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 3
ST-8810
Reminders
The thermometer is not recommended for measuring shiny or polished
metal surfaces such as stainless steel, aluminium, etc. (see chapter
"Emissivity" on page 23).
The thermometer cannot measure through transparent surfaces such as
glass. It will measure the surface temperature of the glass instead.
Steam, dust, smoke, etc., can prevent accurate measurement by
obstructing the thermometer's optics.
Emissivity
Most organic materials and painted or oxidized surfaces (90 % of typical
applications) have an emissivity of 0.95 (pre-set in the unit). Inaccurate read-
ings will result from measuring shiny or polished metal surfaces. To compen-
sate cover the surface to be measured with masking tape or flat black paint.
Allow some time for the tape to reach the same temperature as the material
underneath it. Measure the temperature of the tape or painted surface.
Emissivity values
Substance
Asphalt
Concrete
Cement
Sand
Earth
Water
Ice
Snow
Glass
Ceramic
Marble
Plaster
Mortar
Brick
EN
Thermal
Substance
emissivity
0.90 to 0.98
Cloth (black)
0.94
Human skin
0.96
Lather
0.90
Charcoal (powder)
0.92 to 0.96
Lacquer
0.92 to 0.96
Lacquer (matt)
0.96 to 0.98
Rubber (black)
0.83
Plastic
0.90 to 0.95
Timber
0.90 to 0.94
Paper
0.94
Chromium oxides
0.80 to 0.90
Copper oxides
0.89 to 0.91
Iron oxides
0.93 to 0.96
Textiles
Measurement considerations
Thermal
emissivity
0.98
0.98
0.75 to 0.80
0.96
0.80 to 0.95
0.97
0.94
0.85 to 0.95
0.90
0.70 to 0.94
0.81
0.78
0.78 to 0.82
0.90
23