Herunterladen Inhalt Inhalt Diese Seite drucken

Trouble Shooting Guide; Specifications - Mangar Health Airflo 12 Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Problem
• Compressor does not run when 'up' is
selected.
• Compressor keeps stopping when trying to
inflate the lifting aid.
• Bottom Battery Status Indicator light solid
'Red' and product will not deflate.
• Bottom Battery Status Indicator flashing red
every 2 seconds.
• Bottom Battery status indicator flashing Red
every ½ second.
• Compressor causes electrical interference
with another piece of electrical equipment.
Airflo 12
Weight:
Output pressure:
Batteries:
Battery Life:
Environmental protection:
(Airflo 12 housing)
(Push-button hand control) IP67 (Protected against immersion upto 30 minutes)
Type BF equipment
This equipment has been tested and meets the requirements of current EMC legislation
Charger: Model number - HX1506SR
Input:
Output:
Class II Double insulated equipment.
Healthcare solutions designed for life

Trouble shooting guide

Specifications

3.7Kg (8 lbs)
0.25 bar (3.75 psi)
1 x 12v - 2200mAh Ni-MH battery pack (CD0315)
Greater than 500 charging cycles (1 year depending on usage)
IP41 (Protection from entry by solid objects with a diameter or
thickness greater than 1.0mm & dripping water)
AC 100v - 240v, 50-60Hz, 0.5A Max
DC 15v, 0.35A
Remedy
• Ensure that the air hose from the product
being inflated is connected to the Airflo.
• Check that there are no kinks in the air hose.
• Check that there are no kinks in the air hose.
• Low battery capacity, recharge for a
minimum of 2 hours.
• Internal pump temperature too high - allow a
cool down period of 60-90 minutes.
• Current overload. Check there are no kinks in
in the air hose.
• Move either the compressor or other item
further away from each other until the
interference stops.
Airflo 12