MAX-P
Pediatric O
Sensor
2
Single patient use only
Directions for Use
Indications/Contraindications
The Nellcor® O
single patient use when continuous noninvasive arterial oxygen saturation
and pulse rate monitoring are required for patients weighing between
10 and 50 kg.
The MAX-P is contraindicated for use on patients who exhibit allergic
reactions to the adhesive tape.
Instructions for Use
1. Remove plastic backing from the MAX-P and locate transparent
windows (a) on the adhesive side. Windows cover optical components.
(1)
An index finger is the preferred MAX-P location. Alternatively, apply the
sensor to a small thumb, smaller finger, or great toe.
Note: When selecting a sensor site, priority should be given to an
extremity free of an arterial catheter, blood pressure cuff, or
intravascular infusion line.
2. Orient the MAX-P so the dashed line in the middle of the sensor is
centered on the tip of the digit. Wrap adhesive flaps on non-cable end
around the digit. Note that the cable must be positioned on the top of
the hand or foot. (2)
3. Fold cable end over top of digit so that windows are directly opposite
each other. Wrap adhesive securely around sides of digit. (3)
4. Plug the MAX-P into the oximeter and verify proper operation as
described in the oximeter operator's manual.
Note: If the sensor does not track the pulse reliably, it may be incorrectly
positioned—or the sensor site may be too thick, thin, or deeply pigmented,
or otherwise deeply colored (for example, as a result of externally
applied coloring such as nail polish, dye, or pigmented cream) to permit
appropriate light transmission. If any of these situations occurs, reposition
the sensor or choose an alternate Nellcor sensor for use on a different site.
WARNINGS
1. Do not use the MAX-P or other oximetry sensors during MRI
scanning. Conducted current may cause burns. Also, the MAX-P
M
™ pediatric oxygen sensor, model MAX-P, is indicated for
XI
AX
Latex free
�������
3
Rx ONLY
��