8. Troubleshooting

Problem
Air compressor will not
operate when switched on.
Nebuliser chamber bubbles
or produces little or no mist.
Time taken to nebulise
significantly increases.

9. Technical Specifications

Country
Operating Voltage
Power Consumption
Fuse
Flow Rate
Aerosol output rate
Particle Size (MMAD)
Weight
Dimensions
Type Rating
Sound Pressure Level
Intended Use
Action
If the green light in the mains switch is on, yet the motor does not function,
there is an internal fault. The machine should be returned to the point of
purchase or an authorised person for examination.
If the green light(s) on the switches are off:
1. Check mains lead is securely connected each end.
2. For machines with a 3 pin plug (UK) remove and check the fuse in the plug.
Fuse - 3.15AL 25mm long.
3. Check mains fuse in the control panel. Fuse - T1AL 250V, 20mm long.
4. Disconnect unit from electricity supply. Check mains fuse in the mains
inlet socket.
1. Check inlet filter. Replace if necessary.
2. Wash nebuliser chamber. Replace if problem still occurs.
3. Check driveline tubing is not kinked or leaking. Replace if necessary.
4. Service compressor.
AC2000
UK/EU
220V-240V / 50Hz
62VA (260mA)
1 x T1AL 250V, 20mm long
min flow 7 L/min, max
flow 9 L/min @ 138 kPa*
0.38 ml/min
3.4µ
2.9kg (w/o starter pack)
363 x 230 x 118mm
Class II BF Continuous
60 dbA approx.
Heavy
Econoneb
UK/EU
220V-240V / 50Hz
64VA (270mA)
1 x T1AL 250V, 20mm long
min flow 7 L/min, max
flow 9 L/min @ 138 kPa*
0.39 ml/min
3.4µ
2.7kg
210 x 210 x 185mm
Class II BF Continuous
63 dbA approx.
Heavy
Turboneb 2
UK/EU
230V / 50Hz
68VA (296mA)
1 x T1AL 250V, 20mm long
min flow 9.5 L/min, max
flow 10.5 L/min @ 138 kPa*
0.46 ml/min
3.2µ
2.7kg
210 x 210 x 185mm
Class II BF Continuous
63 dbA approx.
Heavy
en