g) Disposal of the product
Dispose of the product in receptacles for waste
materials of healthcare activities involving
infectious risks.
CAUTION: Not all autoclaves can reach 134° C.
Not all autoclaves draw a pre-vacuum. Please
refer to your autoclave manufacturer for
specific sterilization instructions.

8. 4 DEVICE MAINTENANCE

Important:
- Do not use the device if it appears to be
faulty.
- Do not use an abrasive cleaning agent on the
device.
a) Maintenance
The P5 N
XS control unit, the handpiece
EWTRON
cord and the footswitch must be cleaned and
disinfected on a daily basis (using alcohol or
disinfectants, dental practice wipes, such as
S
™ wipes).
EPTOL
b) Monitoring
The accessories and cords should be regularly
checked for faulty insulation. Replace them if
necessary.
It is important to keep the control unit
ventilation vents clean in order to avoid
abnormal heating.
Check that there is no humidity in the handpiece
connector. If there is, dry it out by wiping, then
blowing with the multifunction syringe.
Note:
In the case of a fault, you are advised to contact
the supplier of your P5 N
unauthorized repairers, who might make your
device dangerous for you and your patients.
IX - MAINTENANCE

9. 1 FUSE REPLACEMENT

Your P5 N
XS is protected by two fuses
EWTRON
located in the mains receptacle (Fig. 2-1).
The replacement procedure is as follows:
- Switch the device off (position O).
- Unplug the power cord from the electric supply.
- Unplug the power cord from the power
receptacle (Fig. 2-1).
- Insert the tip of a flat screwdriver into the
notch above the fuse drawer to release it (Fig.
2-4).
- Remove the old fuses.
- Replace with new fuses of the same type and
rating.
- Push the fuse drawer back into its housing until
it clicks to indicate the correct position.
- Plug the power cord into the receptacle (Fig. 2-
1).
- Plug the power cord into the electric supply.
XS. Do not use
EWTRON
13