Please, comply with the respective valid regulation regarding the return, recy-
cling and disposal of used batteries and accumulators.
If an instrument is not used over an extended time period, the accumulators or
batteries must be removed. Should the instrument be contaminated by leaking
battery cells, the instrument has to be returned for cleaning and inspection
to the factory.

6.4 Fuse Replacement

Prior to fuse replacement, ensure that multimeter is disconnected from
external voltage supply and the other connected instruments (such as
UUT, control instruments, etc.).
Exclusively use fuses of voltage and current values in compliance with
the technical data section.
Using auxiliary fuses, in particular short-circuiting fuse holders is prohibited
and can cause instrument destruction or serious bodily injury of operator.
• Switch of the instrument.
• Loosen the screws on the instrument rear.
• Lift the case cover.
• Remove the defect fuse.
• Insert new fuse.
• Replace the case cover and retighten the screws
Fuse (mA): F200 mA 600 V Ceramic 6.3 x 32 mm
Fuse (A): F10 A 600 V Ceramic 6.3 x 32 mm
Fuse Replacement
35