OPERATING & INSTALLATION
CAUTION:
If the flowrate on the flowmeter ever falls below 0.5L/min,
check tubing or accessories for blocked or kinked tubing or
a defective humidifier bottle.
Meanwhile, humidifier bottle will have air bubble around the sieve core.
Then, oxygen comes from the oxygen outlet.
2. Connect nasal oxygen cannula to oxygen outlet, the other end match the
patient.
Note: The time of absorbing oxygen and the flowrate range according to the
doctor's advices.
*Ⅳ.NEBULIZING OPERATION(8F-3AW, 8F-5AW)
Fill proper medicinal liquid in the nebulizing cup (Please follow the doctor's
advices or don't exceed the maximum reticle of the nebulizing cup).
Upward pull out the nebulizing cover on the nebulizing interface. (Figure 7)
Figure 7: Nebulizing
Connect the air hose to neblizing cup and nebulizing interface, then turn on
the power of the oxygen concentrator, now can begin nebulizing therapy
immediately.
When drug nebulizaing have finished, turn the nebulizing cover right to
nebulizing interface for tightening. If you won't breathe oxygen, please turn
off the oxygen concentration.
Pull out the air hose, pull the mouthpiece, downward pull the nebulizing
cup's cap, empty remained medicinal liquid in the nebulizing cup, then wash
12
Counterclockwis
Nebulizing Cap
Nebulizing Interface
OPERATING & INSTALLATION
the air hose, mouthpiece, nebulizing cup's cap, nebulizing baffle, nebulizing
cup, ripple tube, T-piece, etc , with pure water or immerge them in warm
water for about 15 minutes. In order to wash them healthy, you may add
some vinegar in water. (NOTE: DO NOT cook the accessories above for
wash or wash them with boiled water, in case they'll distort when heating).
After finishing cleaning, must dry all the components before storage.
(Nebulizer installment is shown in figure 8).
NOTE: The timing for using nebulizer must follow the doctor's advices.
Mouthpiece
T-piece
Cap
Baffle
Cup
Figure 8: Nebulizer Components
Ⅴ. ALARM SIGNAL
Concentrator has following failure alarm function:
1) Pressure and cycle failure
2) Compressor failure
3) Low oxygen concentration
Note:All the alarms of the device are low priority.
Ripple tube
Air hose
13