The power cable has not been
vice
correctly inserted into the sock-
ot work
et of the device or the power
socket
The medication has not been in-
serted in the nebulizer
The nebulizer has not been
properly fitted.
vice
ulizes
The nebulizer nozzle is clogged
The air pipe is not correctly con-
nected to the device
The air pipe is bent, damaged or
twisted
The air filter is dirty
pparatus is
The Filter is not correctly
than usual
inserted in its housing
ny
CAUSE
15
SOLUTION
Correctly insert the power cable
in sockets.
Pour the right amount of medi-
cation in the nebulizer
Disassemble the nebulizer and
reassemble it correctly as shown
in the connection diagram on
the cover.
Disassemble the nebulizer, re-
move the nozzle and perform
cleaning operations Failure to
remove medication deposits
from the nebulizer affects its ef-
ficiency and operation.
Strictly comply with the instruc-
tions contained in the CLEAN-
ING, SANITISATION AND DISIN-
FECTION chapter
Ensure the device's accessories
are properly connected to the air
intake (see connection diagram
on the cover).
Unwind the pipe and ensure it is
not crushed or punctured. If nec-
essary replace it.
Replace the filter
Fully and properly insert the Fil-
ter in its housing